Hemodialysis Machine – Common Alarm Signals & Their Management (Set 1)
1. The most common reason for a venous pressure high alarm is:
A. Kinked blood tubing after the dialyzer
B. Arterial needle dislodgement
C. Low blood flow rate
D. Air detected in venous chamber
✅ Correct answer: A
2. What is the appropriate first step when the arterial pressure alarm is low?
A. Flush the dialyzer
B. Check for needle dislodgement or loose connections
C. Increase the blood flow rate
D. Call the nephrologist immediately
✅ Correct answer: B
3. An air bubble detector alarm activates. What is your immediate action?
A. Clamp the arterial line
B. Clamp the venous line and stop the blood pump
C. Increase the heparin dose
D. Adjust the dialysate temperature
✅ Correct answer: B
4. High transmembrane pressure (TMP) alarm may be caused by:
A. Blood leak
B. Dialysate pump failure
C. Clotted dialyzer
D. Air in arterial chamber
✅ Correct answer: C
5. The machine alarm sounds for 'Conductivity Out of Range'. What should you check first?
A. Blood flow rate
B. Dialyzer temperature
C. Dialysate mixing system
D. Heparin line
✅ Correct answer: C
6. A blood leak alarm indicates:
A. Patient has internal bleeding
B. Blood is leaking from the venous chamber
C. Dialyzer membrane is compromised
D. There is air in the arterial line
✅ Correct answer: C
7. If the venous pressure drops suddenly, what should you suspect?
A. Kinked line
B. Needle infiltration or disconnection
C. Clotting in dialyzer
D. Low blood flow rate
✅ Correct answer: B
8. When a 'Temperature Out of Range' alarm occurs, your action should be to:
A. Check and adjust the dialysate heater
B. Increase dialysate flow
C. Lower the blood pump speed
D. Change the dialyzer
✅ Correct answer: A
9. If the UF (ultrafiltration) control alarm is triggered, what is a likely cause?
A. Blood flow is too fast
B. Target UF goal has been exceeded
C. Arterial pressure is too high
D. Patient is hypotensive
✅ Correct answer: B
10. A ‘No Flow Detected’ alarm on the blood pump could be due to:
A. Low dialysate conductivity
B. Blood line disconnection or occlusion
C. Incorrect dialyzer priming
D. Clotted venous chamber
✅ Correct answer: B
11. What does a sudden increase in arterial pressure alarm likely indicate?
A. Venous needle dislodgement
B. Kinked arterial bloodline or needle occlusion
C. Excessive fluid removal
D. Dialysate temperature too high
✅ Correct answer: B
12. A patient moves and the machine suddenly alarms for “venous pressure low.” What should you check first?
A. Dialysate pH
B. Blood leak detector
C. Disconnection or dislodgement of venous needle
D. Dialyzer clotting
✅ Correct answer: C
13. A dialysate flow alarm usually indicates a malfunction in which part of the system?
A. Blood pump
B. Arterial line
C. Dialysate pump or delivery system
D. Air detector
✅ Correct answer: C
14. What action is most appropriate when a 'blood leak' alarm is triggered during treatment?
A. Stop the blood pump and return blood
B. Ignore if patient is stable
C. Flush the dialyzer
D. Increase ultrafiltration
✅ Correct answer: A
15. Which alarm is triggered by increased resistance across the dialyzer membrane?
A. Blood leak
B. Venous pressure high
C. TMP (Transmembrane Pressure) high
D. Air detector
✅ Correct answer: C
16. What is the next step after confirming an actual blood leak?
A. Continue dialysis but slow blood flow
B. Stop dialysis, discard circuit, report incident
C. Recalibrate the blood leak sensor
D. Administer heparin
✅ Correct answer: B
17. The conductivity reading is out of range. What could happen to the patient if dialysis continues?
A. Hyperkalemia
B. Hyponatremia or hypernatremia
C. Hypoglycemia
D. Hyperphosphatemia
✅ Correct answer: B
18. An alarm shows “air detected in venous line.” What is the first thing you must do?
A. Increase blood flow
B. Clamp venous line and stop pump immediately
C. Reset the machine
D. Silence the alarm and observe
✅ Correct answer: B
19. High venous pressure during dialysis could be caused by all EXCEPT:
A. Poor vascular access
B. Clotted dialyzer
C. Air in bloodline
D. Kinked tubing post-dialyzer
✅ Correct answer: C
20. A blood pump stops and displays a “door open” alarm. What should you check?
A. Clamp all lines
B. Check arterial pressure
C. Close and lock the blood pump door properly
D. Re-prime the dialyzer
✅ Correct answer: C
21. What is the primary cause of an arterial pressure too low alarm?
A. Dislodged venous needle
B. Patient hypotension
C. Inadequate blood flow or poor arterial access
D. Dialyzer clotting
✅ Correct answer: C
22. What can cause a high dialysate temperature alarm?
A. UF goal exceeded
B. Malfunctioning heating element
C. Clotted blood tubing
D. High arterial pressure
✅ Correct answer: B
23. If an air detector alarm is activated but no visible air is found, what could be the cause?
A. Loose venous connection
B. Faulty air sensor or damp chamber
C. Low TMP
D. Excessive heparin
✅ Correct answer: B
24. In the event of multiple alarms during dialysis, which principle applies best?
A. Silence all alarms to avoid distraction
B. Stop the machine immediately
C. Prioritize alarms affecting blood circuit safety (air, pressure, blood leak)
D. Call the nephrologist immediately
✅ Correct answer: C
25. A “TMP too low” alarm might indicate:
A. Clotted dialyzer
B. No dialysate flow or disconnected dialysate line
C. Excessive ultrafiltration
D. Venous needle dislodgement
✅ Correct answer: B
26. Which patient condition may falsely trigger high venous pressure alarms?
A. Fever
B. Talking during dialysis
C. Fistula stenosis or obstruction
D. High sodium level
✅ Correct answer: C
27. If the machine displays “heparin pump failure,” what should you do first?
A. Increase blood flow
B. Manually administer heparin after consulting doctor
C. Stop dialysis immediately
D. Restart machine
✅ Correct answer: B
28. Why is it critical to not ignore conductivity or pH alarms?
A. They cause machine damage
B. They can lead to electrolyte imbalances and patient harm
C. They increase UF rate
D. They reduce blood flow
✅ Correct answer: B
29. After resolving an alarm, what is best practice before resuming dialysis?
A. Flush the dialyzer
B. Document alarm only if patient complains
C. Confirm alarm source is resolved and monitor patient
D. Reinsert needles
✅ Correct answer: C
30. What is the safest response to a false air detector alarm that recurs during treatment?
A. Bypass the sensor
B. Manually inspect and clean the air chamber, notify supervisor
C. Deactivate all safety alarms
D. Ignore if blood flow is stable
✅ Correct answer: B
31. During treatment, the “blood pump stopped” alarm is triggered. What should be checked first?
A. Blood leak sensor
B. Heparin pump settings
C. Blood pump door closure and tubing position
D. Dialysate pressure
✅ Correct answer: C
32. What is the risk of overriding or ignoring a “dialysate flow error” alarm?
A. Reduced dialyzer lifespan
B. Inaccurate ultrafiltration rate
C. Inadequate waste removal leading to underdialysis
D. Minor patient discomfort only
✅ Correct answer: C
33. The machine is alarming for “UF control error.” What is the safest immediate action?
A. Restart UF system
B. Stop ultrafiltration and notify biomedical/physician
C. Decrease blood flow rate
D. Reprime the blood circuit
✅ Correct answer: B
34. A recurrent “high TMP” alarm likely indicates:
A. Blood leak detector failure
B. Clotting within the dialyzer membrane
C. Arterial needle dislodgement
D. Venous needle is loose
✅ Correct answer: B
35. If conductivity is too low, the dialysate may cause:
A. Hypernatremia and hypertension
B. Hemolysis and hypotension
C. Fever
D. Hyperkalemia
✅ Correct answer: B
36. Which of the following is not a cause of a “venous pressure high” alarm?
A. Clot in venous line
B. Kinked tubing after dialyzer
C. Disconnected arterial line
D. Poor venous access
✅ Correct answer: C
37. What is the most appropriate response if the “arterial pressure low” alarm occurs and the patient is experiencing signs of shock?
A. Change dialyzer
B. Increase blood pump speed
C. Lower the arterial needle
D. Stop dialysis and initiate emergency response
✅ Correct answer: D
38. During priming, the “air in line” alarm repeatedly triggers. What is the likely cause?
A. Clotted dialyzer
B. Tubing not completely filled
C. Conductivity sensor error
D. UF volume too low
✅ Correct answer: B
39. What is the danger of continuing dialysis after overriding a “temperature too high” alarm?
A. Dialysis is more effective
B. Risk of patient hypothermia
C. Increased clotting in circuit
D. Risk of patient hyperthermia or hemolysis
✅ Correct answer: D
40. If the machine alarms for “blood leak small,” and no blood is visible in the dialysate effluent, what is the best course of action?
A. Continue treatment and monitor
B. Stop treatment and test dialysate for occult blood
C. Ignore the alarm if patient feels well
D. Reduce blood flow to stop leak
✅ Correct answer: B
41. A “blood pump error” message persists even after restarting. What should be your next step?
A. Silence the alarm and continue dialysis
B. Change the entire machine immediately
C. Inspect and re-seat tubing or notify biomedical technician
D. Clamp the venous line and restart heparin pump
✅ Correct answer: C
42. If an air detector alarm is triggered during return of blood, what should you do?
A. Pause and inspect for air, do not resume until line is cleared
B. Ignore if air is not visible
C. Flush air through saline line
D. Continue at a slower pump speed
✅ Correct answer: A
43. The “venous pressure too low” alarm can be triggered if:
A. Arterial needle is dislodged
B. Venous needle is dislodged or disconnected
C. UF goal is too high
D. Heparin dose is too low
✅ Correct answer: B
44. When a “TMP high” alarm is triggered repeatedly despite rinsing and repositioning, what should be done?
A. Increase dialysate flow
B. Replace the dialyzer (possible clotting)
C. Turn off TMP sensor
D. Give more heparin
✅ Correct answer: B
45. What is a major danger if a “blood leak alarm” is ignored?
A. Thrombocytopenia
B. Hemolysis
C. Cross-contamination between blood and dialysate
D. Hypocalcemia
✅ Correct answer: C
46. Which of the following alarms directly indicates a life-threatening risk to the patient?
A. Dialysate flow alarm
B. UF goal alarm
C. Air detector alarm
D. TMP low alarm
✅ Correct answer: C
47. A conductivity too high alarm could result in:
A. Dialysis not removing phosphate
B. Hypertension or fluid overload
C. Hypernatremia, cramping, and headache
D. Dialyzer clotting
✅ Correct answer: C
48. A “heparin pump air bubble” alarm occurs. Your next step is to:
A. Flush the heparin line with saline
B. Restart the blood pump
C. Ignore the alarm
D. Pause dialysis
✅ Correct answer: A
49. Why is it important not to silence alarms without resolving the root cause?
A. Alarms are too sensitive
B. It shortens machine life
C. It violates patient safety protocols and risks patient harm
D. It delays the end of dialysis
✅ Correct answer: C
50. The machine signals “End of Treatment,” but UF goal has not been reached. What action should you take?
A. Restart the entire treatment
B. Recalculate dry weight
C. Check time settings and consult physician if extended time is needed
D. Immediately disconnect patient
✅ Correct answer: C
51. What condition may falsely trigger a “blood leak” alarm even when no leak exists?
A. Bloodline kink
B. UV light sensor error
C. Residual disinfectant or medications (e.g., methylene blue) in blood
D. Low blood flow rate
✅ Correct answer: C
52. What’s the most appropriate initial step when you receive a “machine power failure” alert during treatment?
A. Stop blood pump and wait
B. Immediately clamp bloodlines and prepare for manual blood return
C. Disconnect patient
D. Restart the machine
✅ Correct answer: B
53. What alarm may occur if the dialysate supply is accidentally disconnected during dialysis?
A. UF goal reached
B. Blood pump error
C. Conductivity or flow alarm
D. Venous pressure high
✅ Correct answer: C
54. A persistent “arterial pressure high” alarm can indicate:
A. Kinked tubing before the pump
B. Venous needle infiltration
C. Fistula stenosis or poor blood flow from access
D. Patient hypotension
✅ Correct answer: C
55. During self-care dialysis, the patient reports frequent air alarms. What is a likely cause?
A. Improper priming
B. Low dialysate flow
C. Conductivity mismatch
D. Over-anticoagulation
✅ Correct answer: A
56. Which of the following will NOT cause a “TMP high” alarm?
A. Blood clotting in the dialyzer
B. High ultrafiltration setting
C. Normal saline priming
D. Slow dialysate flow
✅ Correct answer: C
57. What is the most appropriate nursing response to a conductivity alarm that does not resolve after checking connections and restarting dialysate?
A. Continue treatment with lower UF
B. Ignore the alarm if patient is stable
C. Stop dialysis and replace dialysate or machine
D. Increase blood flow to compensate
✅ Correct answer: C
58. “Heparin pump occlusion” alarms may be caused by:
A. Overheating of dialysate
B. Clamped or kinked heparin line
C. Low blood volume
D. Incorrect arterial pressure
✅ Correct answer: B
59. When should the air detector alarm be tested manually?
A. Once every 6 months
B. At the end of each treatment
C. Before every treatment during machine setup
D. After each disinfection cycle
✅ Correct answer: C
60. What is the correct step if multiple alarms (pressure, air, blood leak) occur together and machine cannot be reset?
A. Switch off all alarms
B. Reprogram machine
C. Clamp lines, return blood manually, and change machine
D. Ignore and let alarms settle
✅ Correct answer: C